A Clinical Study of the Effect of Low-Level Laser Treatment on the Clinical Stability of Immediately Loaded Implants in Posterior Maxilla
Alkinan Alashhab1, Isam Alkhoury2, Ghaith Sahtout3*
1 Ph.D. Student at Department of Oral and Maxillofacial Surgery, Faculty of Dental medicine, Damascus University, Syria.
2 Professor at Department of Oral and Maxillofacial Surgery, Faculty of Dental medicine, Damascus University, Syria.
3 Master Student at Department of Orthodontics, Faculty of Dental medicine, Damascus University, Syria.
*Corresponding Author
Ghaith Sahtout,
Master Student at Department of Orthodontics, Faculty of Dental medicine, Damascus University, Syria.
Tel: 00963992635566
E-mail: gsahtout@gmail.com
Received: February 08, 2021; Accepted: March 03, 2021; Published: March 06, 2021
Citation: Alkinan Alashhab, Isam Alkhoury, Ghaith Sahtout. A Clinical Study of the Effect of Low-Level Laser Treatment on the Clinical Stability of Immediately Loaded Implants in Posterior Maxilla. Int J Dentistry Oral Sci. 2021;08(03):1758-1764. doi: dx.doi.org/10.19070/2377-8075-21000374
Copyright: Ghaith Sahtout©2021. This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution and reproduction in any medium, provided the original author and source are credited.
Abstract
Background: Immediate loading of implants became a necessity in order to reduce the period of osseointegration. Photo
biomodulation is a non-invasive treatment that bio-stimulates the new-formed bone on the cellular level in order to improve
the quality of bone around implants.
Objectives: This study aimed to evaluate the efficacy of Low-Level Laser on clinical stability of the immediately loaded implants.
Materials & Methods: 24 Megagen Implants were used in 12 patients in the split-mouth technique. 808 nm GaAlAs Laser
with 4 J power per point was used for two weeks with an application every other day. The two implants in each patient were
loaded immediately (1 week after implantation) by full acrylic crowns. Clinical stability was measured by Mega-ISQ and the
measurements were held in the same operating day, one week after implantation and 3 months after implantation. For statistical
analysis, T-test for independent groups and T-test for paired group were used to determine the differences between the
two groups.
Results: No signifiable differences in ISQ values were noticed between the two groups in the operating day and 3 months
after. While there were signifiable differences one week after implantation.
Conclusions: Within the limitations of this study, using LLLt can improve the bone quality around implants but with no
signifiable differences.
2.Introduction
3.Materials and Methods
4.Discussion
5.Conclusion
6.References
Keywords
Immediate Loading; ISQ; LLLT; Clinical Stability of Implants.
Introduction
Dental implants have high success rates in rehabilitating edentulous
patients if certain conditions are taken into consideration
during treatment. Nevertheless, the prognosis of the implants
remain difficult to predict. Osseointegration depends on several
factors such as the appropriate patient, biologically receptive materials,
the surgical experience of the doctor, and the appropriate
recovery duration [1]. The primary stability depends on the
mesenchymal contact between the bone tissue and the implant
surface immediately after implantation. Secondary stability comes
through the formation of new living bone tissue that fills the
space between the existing alveolar bone and the implant surface
instead of the dead bone [2]. The healing period for implants
varies according to the recommendations of 3-4 months. This
duration increases in the maxillary and the posterior region of the
mandibular due to the increase of spongy bone and it can reach
5-6 months.
In the implants placed in the D4 bone of density, we must add 1-2
months to the previous period, which makes a long duration of
functional loading over the implants [3]. The immediate loading
technique was used in the context of rehabilitating the mouth using
dental implants in order to reduce the treatment duration, as
this technique eliminates the time needed to wait for the recovery
duration and allows the use of temporary compensation immediately
after implant insertion and keeps the implants in a functional
state during the healing period. However, some problems associated with this technology emerged, including the low survival
rate of implants compared to those traditionally loaded immediately
[4]. The implant stability is measured based on a variety
of mechanical, histological, clinical, and chemical methods where
resonant frequency analysis is considered Resonance Frequency
Analysis (RFA) is the non-aggressive clinical method to measure
implant stability performed using Ostell device it gives values
measured by ISQ (Implant Stability Quotient) and it has been
used in many clinical studies.The technique of resonant frequency
analysis is the correlation test of a bone-implant system in which
very small forces are applied by the induction of a transducer.
These forces are applied laterally to the implant and the amount
of movement of the implant is measured. Which practically simulates
the conditions of the occlusion loading of this implant. This
method thus allows giving information about the state of the implant
at any time during the process [5]. Laser therapy is recently
introduced as a non-invasive technique that uses low-power lasers
to improve bone healing. Soft lasers are known in general medicine
for their ability to improve wound healing, collagen synthesis,
and cell multiplication of fibroblasts.Most of the studies focused
on evaluating the bioactivity of these lasers on the multiplication
of osteoclasts, collagen placement, and increased bone morphology
compared with tissues that aren’t exposed to the laser [6].
There are no adequate studies in the medical literature evaluating
the effect of low-power lasers on osseointegrationaround dental
implants, although there are many studies that have suggested a
clinically positive effect of low-powerlaser on implant stability [3],
considering that the bone is of low-density type D3-D4 located
in the upper molar region, it is the bone with the lowest success
rate of dental implant treatment due to the initial lack of stability,
so laser can have a beneficial effect on implant treatment in that
area [7].
Objectives:
Evaluating the efficacy of low-power laser biostimulation on Osseointegration
around implants in the posterior region of the
maxillary and made for immediate loading.
Materials and Methods
The study belongs to experimental studies (a comparison study
between a study group and a control group), the split-mouth design
is applied so that the patient himself was a control sample on
one side and a study sample on the other side.
The study sample
Patients attending the dental implant unit and outpatient clinic of
the department of oral and maxillofacial surgery at the Faculty
of Dentistry at Damascus University, who suffer from symmetric
loss confined to the posterior maxillary teeth and meet the following
inclusion criteria:Adults over 18 years old, posterior symmetric
teeth loss in the maxillary, sufficient bone quantity to receive
the dental implant without the need for any advanced surgical
procedures, absence of any tooth extraction history during the
past six months.
Sample size calculation
The sample size was calculated using the G- power 3.1.9.2 program
based on the data of Torkzaban study 2018 [3], where the
measured ISQ were enteredwitha resonant frequency analyzer device
electronically Osstell that measures the initial stability of the
implant. The result was that a sample size of 24 implants were
applied in 12 patients.
Surgical work
The surgery was performed in the implantation unit at Damascus
University College and dentistry and using the Korean Megagen
implant system. Implants with a united diameter and length (4
mm x 10 mm) were introduced to all patients and at both sides,
according to the following order:
After preparing the surgical field and requesting the patient to
rinse with chlorhexidine solution for 30 seconds, local anesthesia
was performed with filtration from the vestibular and palatine
sides on both ends.
A flap was not raised in this research, but rather a non-flap implant
method was used where starting with the positioning bur in
the implant kit where the implant motor was set at a speed of 800
rpm and a torque of 35 N. We enter after ascertaining the axis of
the bur in both the buccal-lingual and medial-lateral directions
(Figure 1).
The special burs were followed until reaching the last bur for inserting
an implant with a diameter of 4 mm, and this bur in the
system used in the research has a diameter of 3.6 mm, and we
should take care to enter all burs to the full predetermined length,
making sure of the adequacy of irrigation with the syringe and
saline solution, and there is no need to use a counter or any other
bur.
Implants with a diameter of 4 mm and a length of 10 mm were
inserted at the sides, using the surgical handpiece of the implant's
motor, after it was set to the insertion speed of 50 revolutions/
minute and the insertion torque of 35 Newton without using saline
irrigation (Figure 2). The implant was visually confirmed to
reach the level of the apex of the alveolar bone under the periodontal
soft tissue.
Initial implant stability was measured using Megastell from the
vestibular-palatine and medial-lateral aspects.
Install the transfer and prepare the patient in order to begin the
prosthetic procedure.
The compensatory stage
Work in the compensatory phase began immediately after the surgical
work, in the following order:
The transfer impression was installed at both sides, using a closedtray
technique (figure 3).
Condensation silicone, both hard and soft, was used for taking the
imprint, the imprint was taken with a single-stage technique where
the light silicone was injected around the imprint conveyor and
then the tray was filled with the putty silicone.
The imprint was removed from the patient’s mouth and then installed with the special laboratory Analogue implant and was
re-inserted into the imprint according to the perforations on the
conveyor. (figure 4)
Gingival formers were placed on the implants, ensuring the appropriate
size and length for the gingival formers.
Making temporary prostheses using the indirect technique where
imprints are taken and sent to the professional technician to make
temporary acrylic prostheses on the temporary supports of the
Megagen implant system.
The temporary compensation was fabricated on the seventh day
of the surgical operation and it was ensured that there were no occlusal
contacts with occlusion papers (Figure 6). The strut screw
was tightened with a ratchet wrench of the implant system at a
torque of 35 N and the position of the strut screw was closed by
flowable composite.
Laser application
The laser application phase took place in the laser unit at the Faculty
of Dentistry, University of Damascus, and the laser device
used in this research was the Klas DX device - GaAlAs from the
Taiwanese Konftec company (Fig.7), and the following parameters
were adopted:
Capacity of 250 mill watts, wave length 808 nm, application time
16 seconds, total energy of 4 joules in each of the vestibular and
lingual sides.
The laser was used in contact with the tissue after drying and in
the mode of continuous pulse.Use a 0.07 cm crooked head.
In the studied part which determined randomly in each patient,
the gingival tissue was dried, the laser device was set to the previous
barometer, and the device’s head were placed in contact with
the tissue from the vestibular side in the middle of the distance
between the top of the implant and its apex.
After finishing the vestibular side, we move to the palatine side,
where the work is done in the same way. The mucous is dried and
the head is applied in a contact in the middle of the approximate
length of the implant.
The laser was applied in the first two weeks postoperatively, with
48-hour intervals between each application session. (Figure 8)
As for the control sample, the laser was not applied, and the rest
of the procedures were followed, as in the study sample.
Measurement of clinical stability of implants
The clinical stability of the implants was measured using Mega
ISQ device from Megagen and that in:
The same day of the surgical procedure (immediately after implant
insertion).
One week after performing the surgical procedure before installing
the temporary compensation.
After 3 months of the implant procedure, when the temporary
compensation is replaced with the final compensation.
The three measurement times were made in the following order:
The gingival former was removed (at the time of the second
measurement) and the temporary prosthesis with its abutment (at
the time of the third measurement).
Special posts from the Smart Pegs implant company were manually
installed on the implants and were tightened manually until
there were no movement in them (Figure 9).
The Mega ISQ is characterized by its measurement of implant
stability in the vestibular-lingual and medial-lateral directions and
averaging the two measures.
The device head is approximated to the post installed on the implant
without direct contact in one direction until it gives a value
for ISQ and then we adjust the position of the device’s head in
the other direction without contact as well. (Figure 10). The average
value given by the device had been taken and recorded in the
patient data. The work is repeated on the second side in the same
way.
Results
Table (1) shows the descriptive statistics of the clinical stability
values of ISQ implants where:
ISQ 0: The ISQ value measured on the same day of surgery, ISQ
1: The ISQ value measured after one week of surgery, ISQ 2: The
ISQ value measured after 3 months of surgery.
And when performing the Shapieo -Wilk test to determine the
pattern of data distribution, the distribution was found to be normal
(P>0.05) Thus, the value of the parameter tests was chosen.
T-Test was performed for independent samples in order to compare
the two study group and the control group at the three times
the measurement was performed (Table 2).
A t-test was performed for correlated samples in the control
group to compare the ISQ values between each of the two study
times, and statistically significant differences were observed when
comparing each study time table (3).
A t-test was performed for the paired samples in the study group
to compare the ISQ values between each of the two study times,
and statistically significant differences were observed when comparing
each study time Table (4).
Table 2. Shows that there were no statistically significant differences between the two groups in the first time (the same day of surgery) and the third time (3 months after the transplant procedure), while there were statistically significant differences (P <0.05) in the second time (a week after implant).
Discussion
Several clinical and experimental studies have shown promising
results regarding the ability of low-power lasers to improve bone
healing, but they have provided little information on the effect of
these lasers on the bone fusion of dental implants.
GaAlAs laser with a wavelength of 808 nm was chosen due to
its effective effects on both fibroblasts and osteoblasts that have
been proven by several authors in various clinical and laboratory
studies [8, 9].
Lasers in the infrared spectrum have a higher penetration depth
of tissue compared to lasers in the red or blue spectrum in visible
light [10], so bone scanners can absorb laser energy better due to
little water absorption at this wavelength [11].
Khadra has shown that there is no standardized protocol for the
use of lasers for the treatment of dental implants. We have several
options regarding wavelength and laser power when used to
improve bone healing [6].
The Resonnance Frequency Analysis technique was presented by
[12] as a non-aggressive technique for measuring the clinical stability
of implants. This technique is based on the application of
torsion forces that simulate the clinical load on the implants and
its different directions and provides information about the stiffness
of the connection between the bone and the implant, given
that the torsional forces are the most types of stress that are subjected
to dental implants [13].
This technique has been widely used in recent years in order to
measure the initial stability and thus determine the period of time
required for Osseo integration before loading the implant, contributing
to confirm whether sufficient stability has been obtained
in the second surgical phase and thus determining implants with a
high risk of failure [14, 15].
ISQ values gradually decreased in the control group between time
0 (75.25) and time 1 (73.83), then it decreased accordingly after 3
months (68.16), where there were statistically significant differences
in the three times, but in the study group, the ISQ value increased
between time 0 (75) and time 1 (76.5), and then decreased
in time 2 (70.83).
The increase in the value of ISQ in the first week can be explained
by the increase in cell differentiation multiplication and the production
of organic bone matrix around the implants, and that the
power of the laser bio stimulation depends on the physiological
state of the cell at the time of irradiation and on the effectiveness
of the laser during the initial stage of cell proliferation and the
initial differentiation of undifferentiated cells [16].
As for the decrease in the value of ISQ at the third time, it can
be explained that the cells received the laser energy in the initial
stages of the bone restoration and repair process, which accelerated
their differentiation and metabolism during the irradiation
period, and since the application of the laser stopped after 14
days, it can be assumed that the metabolism process is a process.
Related to the energy dose received.
According to Pinheiro, cellular elements in the early stages of
bone healing are more likely to be affected by laser treatment, and
repeated application of the laser is effective when applied in the
cellular stage of bone healing when the number of bone mares is
greater. In later stages, the large number of cells leads to a greater
placement of the bone-like substance that will precipitate out hydroxyapatite,
leading to the maturation of new bone [16].
Scientist Zhou studied the factors that can affect RFA. These factors
include jaw type (mandibular or maxillary), bone quality (D1
-4), patient gender (male or female), implant length and diameter.
Through his study, he revealed that the quality of the bone receiving
the implant is the main factor affecting the ISQ, in contrast
to the hypothesis whichsays that the length and diameter of the
implant took the primary credit for the high value of the clinical
stability of the implants [17].
In our study the bone quality was D2-D3 in all patients, and there
were no differences between the radial bone densities in the two
groups at the time of implantation. This is the probable reason
for the lack of statistical differences between the ISQ values at
different times of measurement.
In the two-way comparison between the two groups, there were
no significant differences between the two groups at the time of
surgery or in the second time after 3 months, while these differences
were found in the first time after a week, meaning that the
low-energy laser showed the greatest effect in the first week after
surgeryand had no effect after 3 months.
The absence of statistically significant differences between the
two groups can be explained by a number of hypotheses, including
that although the laser was applied in one side only without
the other side in the same patient (bisected mouth technique), it
is not certain that the inflammatory effect of the laser is a general
effect. Or is it possible to isolate the induction in the application
area on the osteoblasts, as it has been suggested that low-power
lasers lead to systemic effects in areas far from the place of application
[18], It is also possible that the laser effect was hidden by
the high initial stability of the implants in both groups in addition
to the geometry of the implant. In our study the lowest value of
ISQ was 72.35, which is a large enough value for immediate loading
of implants. For this reason, it is possible that the laser effect
was not the main effective factor in the stability of the implant
as it is one of the influencing factors (initial stability during implantation
- the quality of the bone receiving the implant). Thus,
studies similar to ours - in different conditions of bone quality
(diabetes - osteoporosis - low ISQ values) can further clarify the
potential effect of laser on bone fusion.
The results of our study differed with that of Morales Garcia,
who used GaAlAs laser with a wavelength of 830 nm and did
not find it to have a statistically significant effect. Our study also
disagreed with Mandi who used a 637 nm laser and didn’t notice
the absent effect in improving the ISQ value at all times [7, 19].
The researchers in the two previous studies agreed that the laser
had increased the number of osteogenic cells in the initial stage
of healing, but that this had no effect on the clinical stability.
The results of our study agreed with the Torkzaban study, for
which the ISQ value gradually decreased in both groups and was
the largest decrease after 10 days of implantation, whereas the
ISQ value increased in our study the first week [3].
The results of our study agreed with the results obtained by Tatli,
where he studied the changes in the ISQ value about the immediately
loaded implants and found an increase in the ISQ value in
the first month (in our study, the tip was in the first week), then it
gradually began to decrease later, and this can be explained by the
transformations that occur on the organic template around the
implants after the first month when the reticular bone (lamellar)
begins to turn into dense cortical bone [20].
Conclusions
The effectiveness of the low-power laser in increasing the clinical
stability of implants in the posterior region of the mandibular in
the first week after surgery, while it had no effect on the clinical
stability after 3 months of implantation.
References
- Albrektsson T, Brånemark PI, Hansson HA, Lindström J. Osseointegrated titanium implants. Requirements for ensuring a long-lasting, direct boneto- implant anchorage in man. Acta Orthop Scand. 1981;52(2):155-70.Pubmed PMID: 7246093.
- Tirachaimongkol C, Pothacharoen P, Reichart PA, Khongkhunthian P. Relation between the stability of dental implants and two biological markers during the healing period: a prospective clinical study. Int J Implant Dent. 2016 Dec;2(1):27.Pubmed PMID: 27933572.
- Torkzaban P, Kasraei S, Torabi S, Farhadian M. Low-level laser therapy with 940 nm diode laser on stability of dental implants: a randomized controlled clinical trial. Lasers Med Sci. 2018 Feb;33(2):287-93.
- Gao J, Matsushita Y, Esaki D, Matsuzaki T, Koyano K. Comparative stress analysis of delayed and immediate loading of a single implant in an edentulous maxilla model. J Dent Biomech. 2014 May 14;5:1758736014533982. Pubmed PMID: 25342982.
- Sennerby L. Resonance frequency analysis for implant stability measurements. A review. Integration Diagn Update. 2015;1:1-11.
- Khadra M, Rønold HJ, Lyngstadaas SP, Ellingsen JE, Haanaes HR. Lowlevel laser therapy stimulates bone-implant interaction: an experimental study in rabbits. Clin Oral Implants Res. 2004 Jun;15(3):325-32.Pubmed PMID: 15142095.
- Mandic B, Lazic Z, Markovic A, Mandic B, Mandic M, Djinic A, et al. Influence of postoperative low-level laser therapy on the osseointegration of self-tapping implants in the posterior maxilla: a 6-week split-mouth clinical study. Vojnosanit Pregl. 2015 Mar;72(3):233-40.Pubmed PMID: 25958474.
- Lopes CB, Pinheiro AL, Sathaiah S, Da Silva NS, Salgado MA. Infrared laser photobiomodulation (lambda 830 nm) on bone tissue around dental implants: a Raman spectroscopy and scanning electronic microscopy study in rabbits. Photomed Laser Surg. 2007 Apr;25(2):96-101.Pubmed PMID: 17508844.
- Bouvet-Gerbettaz S, Merigo E, Rocca JP, Carle GF, Rochet N. Effects of low-level laser therapy on proliferation and differentiation of murine bone marrow cells into osteoblasts and osteoclasts. Lasers Surg Med. 2009 Apr;41(4):291-7.
- Welch AJ, Torres JH, Cheong WF. Laser physics and laser-tissue interaction. Texas Hear Inst J. 1989;16(3):141-9.
- Blaya D, Guimarães M, Pozza D, Weber J, de Oliveira MG. Histologic study of the effect of laser therapy on bone repair. J Contemp Dent Pract.2008; 9(6):41-8.
- Meredith N, Alleyne D, Cawley P. Quantitative determination of the stability of the implant-tissue interface using resonance frequency analysis. Clin Oral Implants Res. 1996 Sep;7(3):261-7.
- Sennerby L, Meredith N. Implant stability measurements using resonance frequency analysis: biological and biomechanical aspects and clinical implications. Periodontol 2000. 2008;47:51-66.Pubmed PMID: 18412573.
- Karl M, Graef F, Heckmann S, Krafft T. Parameters of resonance frequency measurement values: a retrospective study of 385 ITI dental implants. Clin Oral Implants Res. 2008 Feb;19(2):214-8.Pubmed PMID: 18067599.
- Aparicio C, Lang NP, Rangert B. Validity and clinical significance of biomechanical testing of implant/bone interface. Clin Oral Implants Res. 2006 Oct;17 Suppl 2:2-7.Pubmed PMID: 16968377.
- Pinheiro AL, Gerbi ME. Photoengineering of bone repair processes. Photomed Laser Surg. 2006 Apr;24(2):169-78.Pubmed PMID: 16706695.
- Zhou W, Han C, Yunming L, Li D, Song Y, Zhao Y. Is the osseointegration of immediately and delayed loaded implants the same?--comparison of the implant stability during a 3-month healing period in a prospective study. Clin Oral Implants Res. 2009 Dec;20(12):1360-6.Pubmed PMID: 19793319.
- Rodrigo SM, Cunha A, Pozza DH, Blaya DS, Moraes JF, Weber JB, et al. Analysis of the systemic effect of red and infrared laser therapy on wound repair. Photomed Laser Surg. 2009 Dec;27(6):929-35.Pubmed PMID: 19708798.
- García-Morales JM, Tortamano-Neto P, Todescan FF, de Andrade JC Jr, Marotti J, Zezell DM. Stability of dental implants after irradiation with an 830-nm low-level laser: a double-blind randomized clinical study. Lasers Med Sci. 2012 Jul;27(4):703-11.Pubmed PMID: 21732113.
- Tatli U, Salimov F, Kürkcü M, Akoglan M, Kurtoglu C. Does cone beam computed tomography-derived bone density give predictable data about stability changes of immediately loaded implants?: A 1-year resonance frequency follow-up study. J Craniofac Surg. 2014 May;25(3):e293-9.Pubmed PMID: 24799099.